Used Agilent G6410B LC-MS/MS
This refurbished Agilent 6410B QQQ LC/MS system can be used for quantitative applications in pharmaceutical ADME/DMPK studies, biomarker validation, clinical research, food safety, forensics, toxicology, and environmental analyses.
This used instrument has been upgraded with 2 turbomolecular pumps for enhanced sensitivity - please see the Limit of Detection and ESI Tune Report.
For method development, the MassHunter software includes an optimizer function, where the analyte can be infused directly into the ionization source or by a series of injections. The software then identifies selected product ions as well as optimizing fragmentor voltage and collision energies from a mix of analytes assayed at once.
6410B Specification
- Source: ESI
- MS/MS sensitivity: Measured as 1 pg of reserpine injected on column, ESI source quantifying on the transition m/z 609 to 195 - Signal-to-noise ratio >500:1
- Mass resolution: Three settings (FWHM) - Unit: 0.4µ; Wide: 1.2µ; Widest: 2.5µ
- Mass accuracy: 0.1µ across the mass range
- Mass stability: <0.1µ in 24 h
- Dynamic range: >6.5 x 106
- Mass range: m/z 5-2,000
- Scan rate: 5,200 µ/sec
- Scan modes: MS scan, MS/MS product ion scan, MRM, MS/MS neutral loss/gain scan and precursor ion scan, SIM
- Minimum MRM dwell time: 2 ms
- MRM transitions: 200 per time segment, thousands possible with multiple segments
- Positive/negative polarity switching: Inter-segment delay of 20 msec
- Collision cell: High-pressure, hexapole with linear acceleration
- Cross-talk: None detectable
- Single point of control: Single-point data system method capability with full control of Agilent 1100 and 1200 Series HPLC systems and 6410 Triple Quadrupole LC/MS
- Time programming: Positive/negative polarity switching in a time segment, Scan and SIM or MRM (plus other modes of data collection), Solvent divert through calibrant delivery system valve, Possible 99-time segments with potential 99 MRMs per-time window
- Ionization source: Electrospray (ESI) with a signal-to-noise ration >500:1
- Autotune: Automated optimization of ion optics and mass axis calibration
- Solvent declustering: Countercurrent gas
- Detector: High-energy conversion dynode and removable electron multiplier horn
- Vacuum system: 2 x triple-stage turbomolecular pump with one mechanical pump
- Dimensions: (W) 1110mm x (D) 660mm x (H) 470mm
- Weight: 107.5 kg
System Includes
- Agilent 6410B LC-MS/MS with ESI probe (option of Multimode ESI/APCI source - please enquire)
- Agilent 1200 LC system including
- Agilent 1200 G1315D DAD
- Agilent 1200 G1316C Column Oven
- Agilent 1200 G1329A Autosampler
- Agilent 1200 G1330B Sample Chiller
- Agilent 1200 G1312B Binary Pump
- Agilent 1200 G1379B Solvent Degasser
- Bottle Tray
- Option of 1290 UHPLC- please enquire
- 1 x Edwards E2M28 Roughing Pump
- PC with MassHunter installed
- TFT Screen
- New Samsung Printer
- UK Installation and warranty as specified
- * Note - 1200 system can be configured to suit your requirements
- ** All modules and MS complete with front panels (some not shown in images).
Warranty Terms & Conditions For Used Equipment
Note: Our warranty terms quoted for this instrument apply to mainland UK customers. For clients outside of this locale, please enquire at point of sale.
Jaytee Biosciences Ltd warranty all reconditioned instruments for a pre-agreed period. This period can be from 30 days to 1 year from the date of installation / delivery of the instrument. Generally newer equipment has a longer warranty than older equipment.
Additional warranty periods can be negotiated at the time of purchase if required.
The list below outlines the items covered by our warranty:
Parts Included In Warranty
- Electronic Assemblies – e.g. PCB’s, Transducers, Interface Cards
- Mechanical Assemblies – e.g. Pump Drive assemblies, XYZ Transport Systems
- Power Supplies – e.g Low and High Voltage Supplies, Lamp Supplies, Cases, Metalwork and Enclosures.
- Major assemblies – PCB’s, Power Supplies, and Displays etc covered for warranty term.
Typical Parts NOT Included In Warranty (LC MSs)
- Wetted Path Components
- Software Updates
- Tubing and Fittings
- Capillaries
- Consumable Parts
- Tools
Enquire
Send us your details and one of our engineers will get back to you with availability and pricing. Alternatively, call us, quoting "st5405", on +44 (0)1227 265333.