Reconditioned Instrument
Waters 2410
This Waters 2410 Refractive Index Detector is designed for high-performance liquid chromatography applications as either a stand-alone unit with an integrator or chart-reader or otherwise as part of a Waters system controller or data system. It features a patented countercurrent heat exchanger and temperature-controlled cell, which allow stable operation under varying conditions.
The flow cell is protected with built-in pressure relief, and the 2410's auto zero and auto purge functions provide automatic operation. With two external column heater controls, auto diagnostics, and battery backup settings, the Waters 2410 offers multiple fail-safe features to ensure versatility and accuracy of results.
Features
- Auto Zero and Auto Purge for Automated Operation
- Built-In Pressure Relief to Protect Flow Cell
- Auto Diagnostics
- Two External Column Heater Controls
- Battery Backup to Retain Parameter Settings When The Detector Is Powered Off Or During Power Interruptions
- Long-Life LED Light Source
| RI Range | 1.00 to 1.75 RIU |
| Measurement Range | 5 x 10-3 RIU maximum FS (SENS = 1, SF = 1) 5 x 10-8 RIU minimum FS (SENS = 1024, SF =100) |
| Flow Rate | 0.03 to 10 mL/min |
| Noise* | ≤ 2.0 x 10-8 RIU, (TC = 1, SF = 80, SENS = 64, THF at 24 °C with restrictor coil). |
| Drift* | 2.5 x 10-7 RIU/hr (static with THF) TC = 1, SF = 20, SENS = 32 |
| Sensitivity Settings | 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024 |
| Time Constant Filter Settings | 0.2, 1, 3, 10 seconds |
| Analog Outputs | -2.0 to +2.0 V |
| Temperature Control | Internal oven: 30 to 50 °C (86 to 122 °F) External column heaters: Ambient to 150 °C (302 °F), steel Ambient to 80 °C (176 °F), plastic |
| * Measurement taken after a minimum 2-hour warmup time or longer for equilibration, depending on column heater. | |
| Integrator Output | |
| Sensitivity Setting | Output (V) |
| 1 | 1 |
| 2 | 2 |
| 4 | 2 |
| 8 | 2 |
| Optical Component Specifications | |
| Light Source: | Light emitting diode (LED) |
| Flow Cell | Fused quartz |
| Cell Volume | 10 µL |
| Pressure Limit | 100 psi |
| Fluidic Path Materials | 316 stainless steel, Dynasil®, Suprasil II®, PTFE, Kalrez®, quartz |
| Environmental Specifications | |
| Operating Temperature | 15 °C to 32.2 °C (59 °F to 90 °F) |
| Operating Humidity | 20 to 80%, noncondensing |
| Shipping and Storage Temperature | -40 °C to 70 °C (-104 °F to 158 °F) |
| Shipping and Storage Humidity | 0% to 90%, noncondensing |
| Dimensions | |
| Height | 8.2 in. (20.8 cm) |
| Length | 19.8 in. (50.3 cm) |
| Width | 11.2 in. (28.4 cm) |
| Weight | 21.7 lbs. (9.7 kg) |
| Electrical Specifications | |
| Line Frequency | 50 Hz, 47 to 53 Hz, 60 Hz, 57 to 63 Hz |
| Fuse Rating | F 2 A, 250 Vac |
| Power Consumption | 140 VA (Nominal) |
| Operating voltage: | 230V |
This detector is in excellent condition - please contact us for full details.
Warranty Terms & Conditions For Used Equipment
Note: Our warranty terms quoted for this instrument apply to mainland UK customers. For clients outside of this locale, please enquire at point of sale.
Jaytee Biosciences Ltd warranty all reconditioned instruments for a pre-agreed period. This period can be from 30 days to 1 year from the date of installation / delivery of the instrument. Generally newer equipment has a longer warranty than older equipment.
Additional warranty periods can be negotiated at the time of purchase if required.
The list below outlines the items covered by our warranty:
Parts Included In Warranty
- Electronic Assemblies – e.g. PCB’s, Transducers, Interface Cards
- Mechanical Assemblies – e.g. Pump Drive assemblies, XYZ Transport Systems
- Power Supplies – e.g Low and High Voltage Supplies, Lamp Supplies, Cases, Metalwork and Enclosures.
- Major assemblies – PCB’s, Power Supplies, and Displays etc covered for warranty term.
Typical Parts NOT Included In Warranty (detectors)
- Cell Assemblies
- Cell Rebuild Kit
- Lamp Assembly
- Wetted Path Components
- Software Updates
- Tubing and Fittings
- Capillaries
- Consumable Parts
- Tools
Enquire
Call us, quoting "st4606 Waters 2410 Refractive Index Detector" on
+44 (0)1227 265333, or alternatively contact us by email.